Use Status Registers to Determine the State of Analyzer Events and Conditions
Questionable Status Integrity Event Condition and
Enable Registers
The Questionable Status Integrity Condition Register continuously monitors the
calibration status of the analyzer. Condition registers are read-only. To query the
condition register, send the command
:STATus:QUEStionable:INTegrity:CONDition? The response will be
the decimal sum of the bits which are set to 1.
The transition filter specifies which types of bit state changes in the condition
register will set corresponding bits in the event register. The changes may be
positive (from 0 to 1) or negative (from 1 to 0). Send the command
:STATus:QUEStionable:INTegrity:NTRansition <num> (negative
transition) or :STATus:QUEStionable:INTegrity:PTRansition <num>
(positive transition) where <num> is the sum of the decimal values of the bits you
want to enable.
The Questionable Status Integrity Event Register latches transition events from the
condition register as specified by the transition filters. Event registers are
destructive read-only. Reading data from an event register will clear the content of
that register. To query the event register, send the command
:STATus:QUEStionable:INTegrity[:EVENt]?
See
"Questionable Status Event Enable Register" on page 77
how to set the summary bit using the event enable register. In this case, use the
command :STATus:QUEStionable:INTegrity:ENABle <num>.