Download Print this page

Display Annotation Title Data - Agilent Technologies EMC Series Programmer's Manual

Spectrum analyzers 9 khz-1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for EMC Series:

Advertisement

Language Reference
DISPlay Subsystem
Factory Preset
and *RST:
The factory default is On. This parameter is persistent, which
means that it retains the setting previously selected, even
through a power cycle.
Front Panel
Access:
System, Time/Date, Time/Date On Off

Display Annotation Title Data

:DISPlay:ANNotation:TITLe:DATA <string>
:DISPlay:ANNotation:TITLe:DATA?
Enters the text that will be displayed in the user title area of the display.
Front Panel
Access:
Display, Title
Display, Title, Change Title
Display, Title, Clear Title
Turn the Entire Display On/Off
:DISPlay:ENABle OFF|ON|0|1
Turns the display on or off. Having the display turned off may increase repetitive
measurement rate.
The following key presses will turn display enable back on:
1. If in local, press any key
2. If in remote, press the local (system) key
3. If in local lockout, no key (the computer must either cancel local lockout, or
re-enable the display)
Factory Preset
and *RST:
On
Turn the Full Screen Display On/Off
:DISPlay:MENU:STATe OFF|ON|0|1
:DISPlay:MENU:STATe?
Turns the full screen display mode on or off. Press
screen mode.
History:
Added with firmware revision A.08.00.
236
to turn off full
System, System
Chapter 5

Advertisement

loading

This manual is also suitable for:

E7404aE7402aE7405aE7401aE7403a