Operating Conditions
Operating Temperature ............................................................... 5 °C to 40 °C
Storage Temperature ................................................................... -20 °C to 70 °C
Humidity ....................................................................................... 80 % maximum for temperatures <31 °C,
Altitude ......................................................................................... <2000 m
Power Requirements ....................................................................... 230 V ac (±10 %), 50/60 Hz, 20 A
Heater Power ................................................................................... 4000 watts at 230 V ac
Over-current Protection
System ......................................................................................... 20 A, 250 V resettable circuit breaker
Main Heater Fuse ........................................................................ F 12 A, 250 V
Zone Heater Fuse ........................................................................ F 12 A, 250 V
Computer Interface .......................................................................... RS-232 and USB
Display ............................................................................................. Monochrome LCD, °C or °F user-selectable
Display Resolution ........................................................................... 0.1 °C or °F
Size (H x W x L) ............................................................................... 400 mm x 337 mm x 700 mm
Net Weight ....................................................................................... 29 kg (without isothermal block)
Isothermal Block (optional)
Isothermal Block Construction ..................................................... Alumina
Block Outer Diameter .................................................................. 37 mm
Block Length ................................................................................ 380 mm
Well Diameter (four wells total) .................................................... 6.7 mm
Well Depth ................................................................................... 365 mm
Net Weight ................................................................................... 0.84 kg (isothermal block only)
78
decreasing linearly to 50 % at 40 °C