SOURce:LCON:DER <n>
Sets the main control derivative time in seconds.
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to
set this value.
Parameters:
n
0.0 to 99.9 seconds.
Example:
SOUR:LCON:DER 5
SOURce:LCON:DER?
Gets the main control derivative time in seconds.
Example:
SOUR:LCON:DER?
Response: 1.5
SOURce:LCON:INT <n>
Sets the main control loop integral time in seconds.
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to
set this value.
Parameter:
n
=
Example:
SOUR:LCON:INIT 10
SOURce:LCON:INT?
Gets the main control loop integral time in seconds.
Example:
SOUR:LCON:INIT?
Response: 20.0
46
The main zone derivative time is 1.5 seconds.
10.0 to 999.9 seconds.
The main control loop integral time is 20.0 seconds.
Note
Note