Fluke 9118A User Manual page 58

Calibration furnace
Table of Contents

Advertisement

9118A
Users Manual
SOURce:PROT:CLEA
Resets the cutout to enable the Product.
Example: SOUR:PROT:CLEA
Response: None
If the furnace goes above the maximum temperature, the cutout circuit will trip. The Product does not heat or cool until the
cutout condition is cleared.
SOURce:PROT:SCUT:LEV <n>
Sets the soft cutout temperature.
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to
set this value.
Parameters:
n
=
Example:
SOUR:PROT:SCUT:LEV 450
SOURce:PROT:SCUT:LEV?
Gets the soft cutout temperature.
Example:
SOUR:PROT:SCUT:LEV?
Response: 1000
SOURce:PROT:TRIP?
Gets the cutout trip state.
Example:
SOUR:PROT:TRIP?
Response: 0
48
25.00 to 1240.00 degrees.
Sets the soft cutout to 450 degrees.
Soft cutout temperature is set to 1000 degrees.
Cutout not tripped (1 = cutout tripped).
Note