Operation - Fluke 9118A User Manual

Calibration furnace
Table of Contents

Advertisement

9118A
Users Manual

Operation

 Warning
For safe operation and maintenance of the
product:
Energize the Product for a 2 hour dry-out
period before use, if the Product was:
In transport
o
In a humid or semi-humid storage
o
environment
Not energized for more than 10 days
o
Remove moisture from the Product before
you connect to mains power. Fluke
Calibration recommends the Product be
put in a low humidity temperature chamber
at 50 °C for 4 hours or more. While the
Product dries, do not assume the Product
complies with all safety standards.
Put temperature probes you remove from
the Product on a heat resistant rack or
surface until they are safe to touch.
Keep a minimum of 45 cm clearance from
the front, left, and right sides. Keep 150 cm
from the top of the Product.
24
In order for the Product to perform within
specification, the open ends of the furnace tube
(both front and rear) must be filled with a
temperature-resistant fiber ceramic insulating
material such as a Superwool Plus Blanket.
Fiber ceramic insulation contains a binding
material that will off-gas when first exposed to high
temperatures producing a harmless odor. Once
the binding material has off-gassed completely,
the odor will no longer persist with subsequent
uses of the furnace.
Warning
To avoid personal injury or skin irritation, wear
disposable latex gloves (or equivalent) when
handling fiber ceramic material.
The front-panel controls let you set the parameters of the
Product and monitor its performance. You can also set the
parameters of the communication interface. Four softkeys
are used to navigate through the menus of the Product.
The menus let you setup the Product and change system
parameters as necessary.