Password Setup Menu; System Information (Info) Menu; Adjust Furnace Temperature; Set The Furnace Temperature With A Preset - Fluke 9118A User Manual

Calibration furnace
Table of Contents

Advertisement

9118A
Users Manual
The LINEFEED parameter turns on and turns off the line
feed character sent across the RS-232 communications
port.

Password Setup Menu

The PASSWORD menu lets you set the user password
and set the level of protection for specified parameters. A
password is necessary to access the password menu.
The USER PASSWORD parameter is used to set the
password for the system that is used to access some
menus. The password is a four-digit number where each
digit is a number from 1 to 9. The default password is
"1234".
The PROTECTION parameter sets the level of password
protection in the Product. If set to HIGH, then a password
is necessary to access the Cutout menu, Ramp/Soak
menu, CALIB (Calibration) menu, and PROG menu. When
set to LOW, only CALIB and PROG menus are password
protected.

System Information (INFO) Menu

The SYSTEM INFO menu is a view only menu. This menu
shows the Product model number, serial number, firmware
version, and the calibration date.

Adjust Furnace Temperature

There are two methods to adjust the furnace temperature.
You can recall a preset temperature or set a temperature
setpoint manually. The term "setpoint" refers to the
temperature the furnace goes to and holds.
32

Set the Furnace Temperature with a Preset

The Product stores eight preset setpoint temperatures.
You can set each preset setpoint as necessary. To set the
setpoint to a preset temperature:
1. Push .
2. Push the softkey below SELECT PRESET.
3. Push  or  to move between presets (1 through 8). In
Figure 21, the temperature of the preset is shown in
the display.
Figure 21. Preset Selection
4. With a preset number highlighted, push .
If the Product is in standby mode, you must push
 again to start the furnace.
Note
gxs111.eps