Fluke 9118A User Manual page 72

Calibration furnace
Table of Contents

Advertisement

9118A
Users Manual
Prepare Test Equipment and Power
On
Set setpoint
Heat On
Test the stabiltiy
Take As Found Data
for stability
Pass/Fail
Pass
Calculate the uniformity
Take As Found Data for
axial uniformity
Fail
Pass/Fail
Calculate the difference
of +/- 30 mm
Set the unifomity align
parameter
62
Fail
Pass
Figure 24. Tube Furnace Calibration Flow Chart
Take As Found Data
for stability
Fail
Pass/Fail
Pass
Take As Found Data for
Axial Uniformity
Fail
Pass/Fail
Calculate the gradient
Set the gradient align
parameter
Take As Found Data
for stability
Fail
Pass/Fail
Pass
Take As Found Data
for Axial Gradient
Fail
Pass/Fail
Pass
Calculate the accuracy
Set the accuracy
parameter
Take As Found Data
for stability
Fail
Pass/Fail
Take As Found Data
for Accuracy
Fail
Pass/Fail
Pass
All
No
Calibration Point
Finished
Yes
Finished
gxs114.eps