SOURce:LCON:PBAN <n>
Sets the main control proportional band in °C which the PID controller uses for main zone control.
Parameters:
n
=
Example:
SOUR:LCON:PBAN 7
SOURce:LCON:PBAN?
Gets the main control proportional band in °C which the PID controller uses for main zone control.
Example:
SOUR:LCON:PBAN?
Response: 1.5
SOURce:LIST:SPO<i> <n>
Sets the temperature value for a preset setpoint in the RAMP/SOAK menu.
Parameters:
i
=
n
=
Example:
SOUR:LIST:SPO6 700.00
SOURce:LIST:SPO<i>?
Gets the temperature value for a preset setpoint in the RAMP/SOAK menu.
Example:
SOUR:LIST:SPO4?
Response: 400.0
SOURce:PROT:HCUT?
Gets the hard cutout temperature setpoint in °C or °F.
Example:
SOUR:PROT:HCUT?
Response: 140
1.0 to 199.9 °C
The main control loop proportional band is 1.5 °C.
Setpoint number (1 to 8)
Setpoint temperature.
Sets setpoint number 6 to 700.00 degrees.
The setpoint temperature for setpoint 4 is 400.0 degrees.
The hard cutout temperature is set to 140 °C.
Calibration Furnace