Fluke 9118A User Manual page 60

Calibration furnace
Table of Contents

Advertisement

9118A
Users Manual
SOURce:SENS:CAL:PAR<x> [n]
Sets a control temperature calibration parameter.
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to
set this value.
Parameters:
x
=
n
=
Example:
SOUR:SENS:CAL:PAR2 0.02
SOURce:SENS:CAL:PAR<x>?
Gets a control temperature calibration parameter.
Example:
SOUR:SENS:CAL:PAR2?
Response: 0.0
SOURce:SENS:CAL:TEMP[n]?
Gets the temperature of a calibration parameter. This temperature is fixed and is in degrees Celsius.
Parameter:
n
=
Example:
SOUR:SENS:CAL:TEMP1?
Response: 300
SOURce:SENS:CAL:SKEW<x> <n>
Sets an axial compensation adjustment parameter.
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to
set this value.
Parameters:
x
=
50
1 to 3 (1 = PAR1 = TEMP 1, 2 = PAR2 = TEMP 2, 3 = PAR3 = TEMP 3, 4 = TEMP 4)
-50.00 to +50.00 (Default = 0.000)
Set the calibration parameter to 0.02 for main control.
The calibration parameter #2 is set to 0.0.
1 to 3 (1 = TEMP1, 2 = TEMP2, 3 = TEMP3). Default = 1
The temperature value of the TEMP1 parameter is 300 °C.
1 to 4 (1 = SKEW1, 2 = SKEW2, 3 = SKEW3, 4 = SKEW4)