SOURce:RATE <n>
Sets the control temperature rate of change (scan rate) in °C or °F per minute.
Parameters:
n
=
Example:
SOUR:RATE 1.26
SOURce:RATE?
Gets the control temperature rate of change (scan rate) in °C or °F per minute.
Example:
SOUR:RATE?
Response: 0.531
SOURce:SENS:CAL:ZONE <x> <n>
Sets the axial gradient adjustment parameter.
This command has password protection. Use the SYST:PASS:CEN command first before you use this command to
set this value.
Parameters:
x
=
n
=
Example:
SOUR:SENS:CAL:ZONE2 0.08
SOURce:SENS:CAL:ZONE <x>?
Gets the axial gradient adjustment parameter.
Example:
SOUR:SENS:CAL:ZONE2?
Response: 0.05
0.10 to 500.00 degrees per minute. Default = 100.00
Sets the scan rate to 1.26 degrees per minute.
The scan rate is set to 0.531 degrees per minute. The response to this command starts out high and
decreases as the temperature gets nearer to the setpoint temperature.
1 to 4 (1 = ZONE1, 2 = ZONE2, 3 = ZONE3, 4 = ZONE4)
-10.0 to 10.0 as a temperature offset in °C.
Axial gradient adjustment is 0.05.
Note
Calibration Furnace
Alphabetic List of Commands
49