3.1 Specifications
Temperature Range
Display Resolution
Stability
Display Accuracy
Well Diameter
Well Depth
Heating Time
Cooling Time
Well to Well
Uniformity
Stabilization
Power
Size
Weight
Fault Protection
Find Quality Products Online at:
Specifications and Environmental
150–1200°C (302–2192°F)
0.1° to 999.9°, 1° above 1000°
±0.5°C
±5.0°C
1.25" (32 mm)
4" (102 mm)
35 minutes to 1200°C
140 minutes with block
±2.5°C (insert "C" at 1200°C)
20 minutes
115 VAC (±10%), 10.5 amps [230 VAC (±10%), 5.2
amps] switchable, 50/60 Hz, 1200 W
12.4" H x 8.2" W x 12.4"D (315 x 208 x 315 mm)
28 lb. (13 kg)
Sensor burnout and short protection, over temperature
thermal cutout
GlobalTestSupply
www.
Chapter 3
Conditions
.com
sales@GlobalTestSupply.com
3-1