9118A
Users Manual
Alphabetic List of Commands
In this section, commands are shown alphabetically. Each command is shown with it long and short form. For each command
there is a description, one or more examples, and a parameter range when applicable.
In this section, square brackets ([ ]) show optional keywords or parameters. Braces ({ }) enclose parameters inside a
command string. Triangle brackets (< >) show that you must replace the parameter with a value. Some commands
use upper and lower case letters. You must use all the uppercase letters in a command. The lowercase letters are
optional. The "|" is used as a separator between optional parameter values in a command.
All commands used with the Product are sequential. Each command is completed before a subsequent command is
processed.
All commands that are not recognized by the Product cause an error message to show in the error queue. A command can be
rejected with an error if the command is misspelled, has incorrect syntax, and/or has invalid parameters.
*CLS
Clears all status registers.
Example:
*CLS
Response: None
*IDN?
Gets manufacturer, model number, serial number and firmware version of the product.
Example:
*IDN?
Response: FLUKE,9118A,0,4.03
*OPT?
Gets product configuration or reference functionality.
Example:
*OPT?
Response: 1
42
Reference hardware is enabled. (0 = disabled)
Note