Connect To Mains Power; How To Turn On The Product - Fluke 9118A User Manual

Calibration furnace
Table of Contents

Advertisement

The fit between the block and the tube is typically
loose in order to accommodate thermal expansion.

Connect to Mains Power

 Warning
For safe operation and maintenance of the
product, remove moisture from the Product
before you connect to mains power. Fluke
Calibration recommends the Product be put in
a low humidity temperature chamber at 50 °C
for 4 hours or more.
To prevent possible electrical shock or
personal injury, connect mains power cord
correctly and make sure the product is
grounded.
Be sure the power cord to the mains supply is wired
correctly. The power cord wires are color-coded as follows:
Black: Line
White: Neutral
Green: Ground
Attach a separate safety ground wire between the
Protective Earth Terminal on the rear panel of the Product
and an earth ground conductor.
Connect the supplied ground wire between the protective
earth terminal on the rear panel and earth ground. Use the
2.5 meter power cord to connect the Product to a 230 V ac
outlet rated for 20 amps.

How to Turn On the Product

Push the "I" side of the power switch, on the front panel of
the Product. The screen shown in Figure 15 shows in the
display while the Product does a self test.
Figure 15. Power-Up Screen
Calibration Furnace
How to Turn On the Product
qrs001.bmp
21