Table 7 is a list of Furnace calibration tests and adjustments.
No.
Test Name
1
Stability Test
2
Axial Uniformity Test
3
Axial Uniformity Adjustment
4
Axial Uniformity Test
5
Accuracy Test
6
Accuracy Adjustment
7
Accuracy Test
Table 7. Furnace Calibration Tests and Adjustments
Test Points (°C)
300, 700, 1000, 1200
300, 700, 1000, 1200
300, 700, 1000, 1200
300, 700, 1000, 1200
300, 700, 1000, 1200
300, 700, 1000, 1200
300, 700, 1000, 1200
Notes
As Found/As Left data to verify the temperature
stability – can be done with other tests at same
nominal setpoints.
As Found data – use data for initial uniformity
adjustment
Adjust the zone and skew parameters to set axial
uniformity to the specification
As Left data to verify the temperature uniformity.
As Found test data – use to adjust the accuracy
Adjusts the TEMP calibration parameters to set the
accuracy to the specification
As Left data to verify the accuracy
Calibration Furnace
Maintenance
63