Fluke 9118A User Manual page 53

Calibration furnace
Table of Contents

Advertisement

OUTPut:STATus <n>
Sets or gets the main heat output on or off.
Parameters:
n
=
Example:
OUTP:STAT 1
OUTPut:STATus?
Gets the main heat output on or off.
Example:
OUTP:STAT?
Response: 0
OUTP[n]:DATA?
Gets the duty cycle of a heater duty cycle in percent.
Parameters:
n
1 to 3 1 = main heater output (default), 2 = zone heater output, 3 = balance heater output.
Example:
OUTP:DATA?
Response: 18.0
Example:
OUTP2:DATA?
Response: 57.0
PROGram:IDEN [quoted string]
Sets the program identifier.
Parameters:
string =
Example:
IDEN? "TEST-1"
ON or 1
Turn on main heat.
OFF or 0
Turn off main heat.
Turn on main heat.
Main heat output is off. (1 = on)
Main heater output duty cycle is 18 %.
Zone heater output duty cycle is 57.0 %.
1 to 12 characters (0 - 9, A – Z, '-'). Default = "0"
Calibration Furnace
Alphabetic List of Commands
43