Fluke 9118A User Manual page 61

Calibration furnace
Table of Contents

Advertisement

n
=
Example:
SOUR:SENS:CAL:SKEW2 0.08
SOURce:SENS:CAL:SKEW<x>?
Gets an axial compensation adjustment parameter.
Example:
SOUR:SENS:CAL:SKEW2?
Response: 0.05
SYST:CONF:BLC [n]
Sets the furnace work mode with the isothermal block inside or without the isothermal block inside.
Parameters:
n
=0 determines if the furnace will work without the isothermal block inside
=1 determines if the furnace will work with the isothermal block inside
Example:
SYST:CONF:BLC 1 Determines if the furnace is to work with the isothermal block inside.
SYST:CONF:BLC?
Gets the furnace work mode with the isothermal block inside or without the isothermal block inside.
Example:
SYST:CONF:BLC?
Response: 0 The furnace works without the isothermal block inside
SOURce:SENS:DATA? [heater part]
Gets the control temperature of the parts of the furnace.
Parameters:
heater part
Example:
SOUR:SENS:DATA? MAIN
-10.00 to 10.00
The axial compensation for SKEW2 is 0.05 of the main heater power.
MAIN – gets main heater temperature (Default)
ZONE – gets zone heater temperature
FULL – gets main and zone heater temperature
Calibration Furnace
Alphabetic List of Commands
51