Maintenance

 Warning
To prevent possible electrical shock, fire, or
personal injury:
Do not operate the Product with covers
removed or the case open. Hazardous
voltage exposure is possible.
Use only specified replacement parts.
Have an approved technician repair the
Product.
 Caution
To prevent possible damage to the Product:
Do not use fluids to clean the well.
Do not use harsh chemicals on the Product
surfaces.
Do an over-temperature cutout test a
minimum of one time in a 6 month period.
Make sure the well is clean and clear of
unwanted material.
Use the National Safety Council
decontamination steps when hazardous
material is spilt on or in the Product.
Never introduce any foreign material into
the probe hole of the Isothermal Block.
Fluids, etc., can leak into the instrument
causing damage.
Do not lift or move this instrument with the
isothermal block in place. The isothermal
block and rear plug may fall out of the
Product.
Fluke Calibration recommends that you do an axial
uniformity test when you receive the Product and at
6 month intervals. See the Furnace Calibration section.
Calibration Furnace
Maintenance
59