Fluke 9118A User Manual page 54

Calibration furnace
Table of Contents

Advertisement

9118A
Users Manual
PROGram:IDEN? <quoted string>
Gets the program identifier.
Example:
IDEN?
Response: TEST-1
PROGram:SEQ:CAT?
Gets a parameters list for ramp and soak tests.
Example:
PROG:SEQ:CAT?
Response: "SPOn","DWELL","DIR","POIN","CYCL"
PROGram:SEQ:PAR <Par>[, n]
Set a parameter in the ramp/soak test.
Parameters:
Par
=
To make a ramp/soak test with 3 setpoint temperatures of 450 °, 1145 °, and 825 ° that soak for 20 minutes:
Example:
PROG:SEQ:PAR POIN,3
PROG:SEQ:PAR SPO1,450.00
PROG:SEQ:PAR SPO2,1145.00
PROG:SET:PAR SPO3,825
PROG:SET:PAR DWEL,20
44
x = 1 to 8 setpoint number. Default = 1. n = 300.00 to 1200.00. Default = 300.00.
SPO[x]
1 to 100 seconds of soak time. Default = 15.
DWEL
1 to 8 setpoints in the ramp/soak test. Default = 8.
POIN
1 to 999 times the ramp/soak test is to be repeated. Default = 1.
CYCL
0 or 1. Sets the direction of how the setpoints will be done in the test. See DIRECTION
DIR
in the Ramp/Soak Menu section. 0 = Up. 1 = Up then Down. Default = 0.
Sets three temperature setpoints for the test.
Sets setpoint #1 to 450.00 degrees.
Sets setpoint #2 to 1145.00 degrees
Sets setpoint #3 to 825.00 degrees
Sets soak time to 20 minutes.