Calibration Procedure For 9118A-Itb (With Isothermal Block); Calibration Equipment Setup - Fluke 9118A User Manual

Calibration furnace
Table of Contents

Advertisement

9118A
Users Manual
Calibration procedure for 9118A-ITB (with
Isothermal Block)
Caution
The furnace can be used with or without an
isothermal block installed. Unique calibration
parameters are necessary for each mode of
operation. To prevent damage to the Product,
before Product use, select the appropriate
mode of operation in the calibration menu (see
Furnace Setup Menu). When the mode of
operation is changed, a calibration for the new
mode of operation may be required.
Figure 27 shows a brief overview of the calibration
procedure with a thermo block.
70

Calibration Equipment Setup

1. Mark scales on TC1 and TC2 as shown in Figure 26.
Scale interval is 10 mm.
2. Connect equipment as shown in Figure 25.
3. Put TC1 and TC2 in the position of 0 and 60 as shown
in Figure 26.
In the steps that follow, start the calibration
procedure at the lowest temperature (300
move up the range to 1200
one of the setpoint temperatures in Table 15.
Note
°
C) and
°
C. Set the Product to