Fluke 9118A User Manual page 82

Calibration furnace
Table of Contents

Advertisement

9118A
Users Manual
Prepare Test Equipment and Power
On
Set setpoint
Heat On
Test the stabiltiy
Take As Found Data
for stability
Pass/Fail
Pass
Calculate the accuracy
Pass/Fail
Fail
Set the accuracy
align parameter
Take As Found Data
for stability
Pass/Fail
Pass
Take As Found Data
for accuracy
72
Pass
Fail
Fail
Figure 27. Calibration Procedure for Furnace with Thermo Block
Calculate the Axial
uniformaty
Take As Found Data for
Axial Uniformity
Pass/Fail
Fail
Calculate the Slope of
0 mm and 60 mm
Slope <0.1 C
Pass
Set the uniformity align
parameter
Take As Found Data
for stability
Fail
Pass/Fail
No
Calculate the Zone
Set the Zone align
parameter
Yes
Take As Found Data
for stability
Pass/Fail
Pass
Take As Found axial
uniformity
Pass
All
Calibration Point
Finished
Yes
Finished
Fail
gxs135.eps