Fluke 9118A User Manual page 66

Calibration furnace
Table of Contents

Advertisement

9118A
Users Manual
SYSTem:LANG?
Gets the display language.
Example:
SYST:LANG?
Response: 1
SYSTem:PASS:CDIS
Disable access to commands with password protection.
Example:
SYST:PASS:CDIS
This command has no response.
SYSTem:PASS:CEN [n]
Enables access to commands with password protection.
Parameters:
n
=
Example:
SYST:PASS:CEN 1045
This command has no response.
SYSTem:PASS:CEN:STAT?
Gets the access state of password protected commands.
Example:
SYST:PASS:CEN:STAT?
Response: 0
SYSTem:PASS:NEW <n>|DEF
Sets the system password.
56
The display language is set to English.
0000 to 9999. If the n parameter is left out, then the default password of 1234 is used.
Disables password protection on commands where a password is necessary to set
them. Password protection remains disabled until SYST:PASS:CDIS is used or you
turn off power to the Product.
Cannot change commands with password protection. A response of 1 means password commands
can be set by the applicable command. Use the SYST:PASS:CEN command to set the password
protection parameter.

Advertisement

Table of Contents
loading

Table of Contents