PID Setup Menu

The PID SETUP menu is used to set controller
parameters.
The TEMP PB parameter is the main proportional band
and the gain that the proportional-integral-derivative (PID)
controller uses for control.
The TEMP INT parameter lets you set the time the
controller uses for control. The time is in seconds.
The TEMP DER parameter lets you set the derivative time
the controller uses for control. The time is in seconds.

Furnace Setup Menu

To operate the Product with or without an isothermal
block.
1. From the calibration menu, push the Furnace Setup
softkey.
2. Enter the password.
3. Push .
4. Push  or  to scroll through the display status of
"YES" and "NO". Select "YES" to set the Product to
work with the isothermal block installed. Select "NO" to
set the Product to work without the isothermal block
installed.
5. Push  to save the selected mode.
Caution
The Product can be used with or without an
isothermal block installed. To prevent damage
to the Product, unique calibration parameters
are necessary for each mode of operation.
Before Product use, select the appropriate
mode of operation in the calibration menu.
When the mode of operation is changed,
calibration for the new mode of operation may
be required.

System Menu

The SYSTEM menu lets you get access to the display
menu to set the language, set the type of decimal
character, and turn off and turn on key audio. The
communication setup parameters, password, and system
information are also available through this menu.

Display (DISP) Menu

The LANGUAGE parameter is used to set the language
used in the display.
The DECIMAL parameter is used to set the decimal
character to a period or a comma.
The KEY AUDIO parameter is used to turn off and turn on
the audio feedback when a key is pushed.

Communications (COMM) Setup Menu

The BAUD RATE parameter lets you set the baud rate for
the RS-232 communications port.
Calibration Furnace
Operation
31