Table 20 shows the value for B at each temperature
setpoint.
Table 20. Zone Correction Values For Isothermal Block
Calibration Point
300
700
1000
1200
15. Put the ZONE
(new) value into the Product through the
n
front panel or RS-232 port. Table 21 is a list of
temperature setpoints and the RS-232 command and
front-panel name for the ZONE parameters.
B Value
30
30
10
10
Calibration Procedure for Tube Furnace
Table 21. Zone Parameter for Isothermal Block
Setpoint
(°C)
RS-232 Command
300
SOUR:SENS:CAL:ZONE1
700
SOUR:SENS:CAL: ZONE 2
1000
SOUR:SENS:CAL: ZONE 3
1200
SOUR:SENS:CAL: ZONE 4
16. Let the Product soak for 60 minutes for the calibration
points of 700 °C, 1000 °C, and 1200 °C. Let the
Product soak for 120 minutes for calibration points of
300 °C after you put in the ZONE value.
17. Do steps 1 through 5 again to make sure axial
uniformity is equal to or less than the specification or
requirement. If the Product does not satisfy the
specification or requirement, do steps 6 through 16
again with the new axial uniformity values.
To complete the calibration of the Product, do a stability
test, axial uniformity test, and accuracy test to make sure
the Product is equal to or less than the specifications for
each setpoint.
Calibration Furnace
Parameter
Front Panel
SKEW (300)
SKEW (700)
SKEW (1000)
SKEW (1200)