AMI Phosphate-II
Installation
3.5.
3.6.
3.6.1
Terminals
1)
NC
10/11
Normally
Closed
NO
12/11
Normally
Open
A-96.250.641 / 290121
Input
Note: Use only potential-free (dry) contacts.
The total resistance (sum of cable resistance and resistance of
the relay contact) must be less than 50 Ω.
If signal output is set to hold, measurement is interrupted if input is
active.
For programming see menu
Relay Contacts
Alarm Relay
Note: Max. load 1 AT / 250 VAC
Alarm output for system errors.
Error codes see
Error List, p. 55
Programming see menu
Note: With certain alarms and certain settings of the AMI
transmitter the alarm relay does not switch. The error, however, is
shown on the display.
Description
Active (opened) during normal
operation.
Inactive (closed) on error and
loss of power.
Active (closed) during normal
operation.
Inactive (opened) on error and
loss of power.
1) usual use
5.3.4, p. 79
5.3.1, p. 75
Relay connection
1)
0V
0V
11
10
12
11
10
12
23