Input; Relay Contacts; Alarm Relay; Relay 1 And 2 - Swann AMU-II Oxytrace Operator's Manual

Table of Contents

Advertisement

AMU-II Oxytrace
Installation
3.6.
3.7.
3.7.1
3.7.2
A-96.150.541 / 111223

Input

Note: Use only potential-free (dry) contacts.
Terminals 16/8
For programming see

Relay Contacts

Alarm Relay

Note: Max. load 100 mA/50 V
Alarm output for system errors. For error codes see
Terminals
NC
9/10
Normally
Closed
NO
1/2
Normally
Open

Relay 1 and 2

Note: Max. load 100 mA/50 V
Terminals 3/4: Relay 1
Terminals 5/6: Relay 2
For programming see
Installation menu.
Program List and Explanations, p.
Description
Active (opened) during normal
operation.
Inactive (closed) on error and loss of
power.
Active (closed) during normal
operation.
Inactive (opened) on error and loss of
power.
Program List and Explanations, p.
53.
Error List, p.
44.
53,
19