Section 9
User Settings

Use of Standby Mode

Selecting the
Setpoint
9-4
Nanopure TOC - Analytical
At the end of the work day, place the Nanopure TOC - Analytical in
Standby mode for the night. To place the unit in Standby mode:
1. From the (Idle) mode, when display reads, "Nanopure (Idle) xx:xx:xx
xx/xx/xx," press STANDBY.
—OR—
2. From the xx.x MΩ-cm or x.xx uS/cm operational display, Press
STANDBY.
After STANDBY is pressed, the display will read "Standby" and the green
LED will slowly blink. For 10 minutes of every hour while the pump is
energized, the display will read "Standby (Recirculating) xx minutes." Press
START/STOP to return to the recirculating mode.
Note When in Standby mode, press START/STOP to continue normal
operation.
s
Note If you have installed an Accudispense volumetric remote dispenser, its
"Purity Above Setpoint" LED will slowly blink when the water system is in
Standby Mode.
s
Note Refer to Auto Standby section to have the unit automatically enter
Standby Mode each day at a predetermined time chosen by the user.
The Nanopure TOC - Analytical electronics include a user programmable
set point which alerts you when water quality falls below the programmed
set point. The set point is user selectable from 1-18 MΩ-cm (factory set at
10 MΩ-cm.).
1. From the (Idle) mode, when display reads, "Nanopure (Idle) xx:xx:xx
xx/xx/xx," press the UP or DOWN arrow until display reads,
"Advanced Menu?"
2. Press ENTER.
3. Proceed through the options by pressing ENTER until "Adjust Set
Point" is displayed.
4. Press the UP or DOWN arrow to choose "Yes."
5. Press ENTER.
s