The Start/Stop parameter options are Proximity Start, Remote Trigger, and Easy Trigger.
There is also a field to set the maximum time for sample analysis before the analysis stops.
Select the Proximity Start checkbox to use the Proximity Start parameters. Using Proximity
Start, once the reading has been started, release of the Proximity Button will immediately stop
the analysis. You cannot use Proximity Start with Easy Trigger.
Select the Remote Trigger checkbox to use the Remote Trigger parameters. Remote Tigger is
used with the Extend-a-Pole accessory to control the analysis. With the Extend-a-Pole's input
cable connected to the analyzer's Remote Trigger port, you can initiate and stop analysis
remotely from the Extend-a-Pole's handle trigger. You can use Remote Trigger with either
Proximity Start or with Easy Trigger.
Select the Easy Trigger checkbox to use the Easy Trigger parameters. Easy Trigger uses a single
press and release of the trigger to initiate analysis, and a second press to stop analysis. Selecting
Easy Trigger will immediately disable Proximity Start.
Figure 57.
The Max Test Time Editor