Sorting the Custom Element Display

Third Beep

Beep on Grade Match

Sorting the Custom Element Display

Display Options

66
Niton XL2 Analyzer User's Guide
This option allows you to change the seconds of delay before the Third Beep. Select the screen
button labeled with the number of seconds of delay for the Third Beep. The Beep Three Time
editor will open. Clear the current number of seconds with the "C" button, then select the E
button to enter the information.
Selecting this option will enable a special beep when the reading chemistry matches an alloy
grade, and put a check mark in the box. Selecting the box again will remove the check mark
and turn the beep off
Select the Custom Element Display icon to configure sorting criteria used for analysis display.
Select the mode you wish to change, then selecting the Custom Element Display icon opens
up the Custom Element Display Screen.
On the left of the display are elements, each with its currently selected display option beside it
to the right. The element list is ranked by importance, with the most important element on
top, and each one lower down of less importance than the one above it.
By selecting an element and using the arrow buttons to the right of the list, you can change its
ranking. Use the Up Button to move an element one rank closer to the top with each click.
Use the Dn Arrow Button to move an element one rank closer to the bottom with each click.
The Display Options Drop Down Menu allows you to change the display status of any
element to one of three states:
• Normal - The standard state. Element displays only when the elemental value is
greater than the limit of detection.
• Always - Always display the results for this element. Use this state for elements critical
to all of your analyses.
• Never - Never display the results for this element. Use this state for elements which
are unimportant to your work. This makes your instrument display less complex.
Select the element you want to change, then select the menu option corresponding to your
choice of display status. The currently selected element is displayed in white on black.
Select the Save Button to save your current status as the new default. Select the Reset button
to reset the settings to the previously saved state. Select the Close button to exit the screen
without saving.