• When the bracket is clean, remove the backing from the Measurement Window. Place the
window on the Bracket gently. Make sure the opaque portions of the window do not
intrude over the large measurement hole in the Bracket.
Figure 5.
Removing the Backing from Prolene Window (Left) and Applying Window to Bracket (Right)
CAUTION
Do not use fingers to press window into place! Use a smooth, hard surface such as back
of tweezers.
• Replace Window Bracket on the front of your analyzer, then insert screws.
Figure 4.

Prolene Window PN 187-1454

Figure 6.

Measurement Window Replaced

8
Learning More, Service, and Support
Replacing the Measurement Window
Niton XL2 Analyzer User's Guide
131