Table 5 - Scatter Measurements off various substrates - Dose Rates in
Table 5. Niton XL2 Radiation Profile - Scatter Measurements -
kV
uA
Range
15
80
Low
15
80
Low
15
80
Low
15
80
Low
20
80
Low
20
80
Low
20
80
Low
20
80
Low
45
44
Main
45
44
Main
45
44
Main
45
44
Main
Thermo Scientific
Substrate
Max @ 5cm
Aluminum
<0.1
Stainless
<0.1
Plastic
<0.1
Soil
<0.1
Aluminum
<0.1
Stainless
<0.1
Plastic
30
Soil
Aluminum
12
Stainless
16
Plastic
190
Soil
Notes:
Scatter measurements were taken at a radius of 5 or 30 cm around the nose of the analyzer
with the highest scatter dose rate being recorded.
µ
Max @ 30 cm
<0.1
<0.1
<0.1
<0.1
<0.1
<0.1
<0.1
<0.1
<0.1
0.17
<0.1
12
20
0.50
2
Safely and Effectively Using Your Analyzer
Sv/hr
µ
Sv/hr
Max @ Trigger
<0.1
<0.1
<0.1
<0.1
<0.1
<0.1
<0.1
<0.1
0.1
<0.1
1.5
0.25
Niton XL2 Analyzer User's Guide