Report Settings

RoHS Option

Detection Option

Consumer Products Option

Max Measure Time

Minumum Test Time

Thermo Scientific
Under Electronics Metals, Plastics, and Test All Modes, A field called Report Settings is
available. Selecting the triangle next to the Report Settings Field will open a pop up menu
allowing you to choose between the three Report Settings Modes. Select the mode you wish to
edit.
Changing the settings for one analysis mode will not affect the settings for other modes, and
the configurations can be saved independently.
When the RoHS Option is selected, clicking on the Pass and Fail values works as it does in
any other Mode.
When the Detection Option is selected, Selecting the Pass/Fail field for that element acts as an
On/Off Toggle, which will switch Pass/Fail mode between On and Off for the selected
element. Selecting it again will reverse the toggle.
When the Consumer Products Option is selected, clicking on the Pass and Fail values works
as it does in any other Mode. In addition, the total of Cl+Br is also calculated and used for
Pass/Fail Testing.
Under the Method Setup -> Measurement Parameters option is a field called Max Measure
Time. Here you can set up the maximum time your analyzer will continue to analyze the
sample. Select the Max Measure Time field, and a Virtual Numeric Keypad will pop up,
allowing you to input a new Maximum Measurement Time in seconds. The default Max
Measure Time is set to 300 seconds.
Under the Method Setup -> Consumer Goods option is a field called Minimum Test Time.
Here you can set up the minimum time your analyzer will continue to analyze the sample
when using the Detection Option only. Select the Minimum Test Time field, and a Virtual
Numeric Keypad will pop up, allowing you to input a new Minimum Test Time in seconds.
The default Minimum Test Time is set to 60 seconds.
5
Common Operations
Max Measure Time
Niton XL2 Analyzer User's Guide
67