Calibrating the Touch Screen
Thermo Scientific
Figure 19. Initiating Touch Screen Calibration
Select the Calibrate Touch Screen button from the System Screen to re-calibrate the analyzer's
touch screen display. This procedure establishes the display boundaries for the touch screen
interface.
1. Select the Touch Screen icon.
2. The display will show a message asking you to confirm whether or not you want to
calibrate your Touch Screen. Select the Yes button.
3. The display will show the message: "Calibrate Touch Screen". There will be a small cross in
the upper left-hand corner of the display.
4. Tap on this cross with the stylus, and the cross will disappear and reappear in the upper
right-hand corner of the screen.
5. Tap on the cross again, and it will reappear in the lower right-hand corner of the screen.
6. Tap on the cross again and it will reappear in the lower left-hand corner of the screen.
7. Tap on the cross once more, and you will be presented with a Confirmation Screen.
8. Select the Yes Button to confirm that the parameters are good. Select the No Button to start
the process again.
5
Common Operations
Calibrating the Touch Screen
Niton XL2 Analyzer User's Guide
81