Niton Xl2 Goldd Radiation Profile - In Beam Measurements - Rem/Hr; Niton Xl2 Goldd Radiation Profile - In Beam Measurements - Msv/Hr - Thermo Scientific XL2 User Manual

Table of Contents

Advertisement

2
Using Your Analyzer
Safely and Effectively Using Your Analyzer
Table 10 - Niton XL2 GOLDD Radiation Profile - In Beam Measurements - Rem/hr
Table 10. Niton XL2 GOLDD Radiation Profile - In Beam Measurements - Rem/hr
kV
uA
Range
8
100 Light
45
44.4 Main
Table 11 - Niton XL2 GOLDD Radiation Profile - In Beam Measurements - mSv/hr
Table 11. Niton XL2 GOLDD Radiation Profile - In Beam Measurements - mSv/hr
kV
uA
Range
8
100 Light
45
44.4 Main
16
Niton XL2 Analyzer User's Guide
Contact Deep
0.01
39
Contact Deep
0.1
390
Notes:
In beam dose rates were measured using optically stimulated luminescent (OSL) dosimeters.
Reported results are based on measurement results that have been reduced to 2 significant
digits by rounding up. For example, a measurement result of 1441 would be reported as
1500.
Contact Shallow
5cm Deep
980
0.002
130
8.1
Contact Shallow
5cm Deep
9800
0.02
1300
81
30 cm Deep
<0.001
0.51
30 cm Deep
<0.01
5.1
Thermo Scientific