Learning More, Service, and Support

Replacing the Measurement Window

This section of the Resource guide is about getting the most out of your
analyzer. We cover troubleshooting your analyzer by using the Specs
screen. We also cover advanced topics like setting thresholds, using the
Tools menu, correcting for light elements in the sample composition,
setting up pass/fail analysis, changing safety and start/stop parameters,
and many other special situations you may need. We have also included
a number of documents for reference, so you can learn more about XRF
analysis if you are so inclined.
WARNING
Before you begin, remove the battery from your analyzer!
• Remove the two screws holding the Measurement Window Bracket to
the nose of your analyzer.
Figure 1.
Removing the Window Bracket Screws
• Remove the Measurement Window Bracket from the analyzer, and turn
it over, exposing the back with seal and Measurement Window.
• Remove the old Measurement Window from the bracket.
• Clean the Window area thoroughly, using a clean, guaranteed lint-free
cloth and isopropyl alcohol.
129