Using Your Analyzer
Proper and Improper Operation
20
Niton XL2 Analyzer User's Guide
1. Stop using the instrument
2. Remove the battery. The x-ray tube can not produce radiation when the battery is
disconnected. The instrument is now safe to handle.
3. Place the instrument securely in the holster.
4. Place the instrument in the carrying case that came with the instrument.
5. Notify your Radiation Safety Officer (RSO) or the equivalent responsible individual at
your company or institution immediately.
6. You or your RSO should call Thermo Fisher Scientific at one of their contact numbers
listed below for additional instructions and guidance.

Major Damage

If the instrument is severely damaged:
1. Perform the same steps as described above for minor damage. There will be no radiation
hazard as long as the battery is removed from the instrument.
2. Place all components in a plastic bag and contact Thermo Fisher Scientific.