Diagnostics

Thermo Scientific
Select the "->" Screen Button to load the Diagnostics Screen. The Diagnostics Screen
shows Detector Temperature, Bias, Cooler Voltage, SubBias, Energy Scale, and
Temperature in C and F scales.
The Diagnostics Screen can be of great utility in assuring proper operation of your
analyzer.
Figure 10.

The Diagnostics Screen

The proper ranges of operational values on the Diagnostics Screen follow.
Det Temp:
Detector Temperature should be within this range:
- 25 + or - 5 degrees
Bias:
Bias should be within this range:
175 + or - 10
VCool:
VCool will vary with the ambient temperature.
8
Learning More, Service, and Support
Tips and Troubleshooting
Niton XL2 Analyzer User's Guide
135