FPGA

Factory QC

Energy Cal

Battery

System Check Reminder
134
Niton XL2 Analyzer User's Guide
This is the currently loaded FPGA software version, which should be reported to Service
if there is any problem. FPGA versions are always a four digit number. Any other number
of digits may be a sign of a problem in the FPGA code.
This is the date that the machine was QCed at the factory.
This line notes the last time a System Check was performed.
This line gives the proportional charge remaining to the battery.
Select the OFF Screen Button after "System Check Reminder" to set a reminder to
calibrate your analyzer. Selecting the button will open the Cal. Reminder Editor. Select
the number of days you want between reminders with the numeric keys. Of the other
screen buttons, C = Clear All, E = Enter, and OFF shuts off the Reminder Function.
Selecting the E button will enter the current value as the reminder interval and return to
the Specs Screen.
Figure 9.
Cal Reminder Editor Screen