Connectivity
Downloading Data
Incorrect Data in reading # XXX.
Reading will be skipped.
Error code:
BOUNDARY_ERROR3.
WARNING: 115200 baud rate not
supported.
SH4 Successfully Communicating
Result: SUCCESS

Live Download (Automatic Save)

122
Niton XL2 Analyzer User's Guide
Table 4: Error Messages while Downloading
This indicates a version mismatch between your
instrument code and the NDT code running on your
computer. Use a version of NDT that matches the
ver?sion number of the software on your analyzer.
This indicates a potential problem. Test the serial port.
If there is a problem connecting, switch baud rate on
both the analyzer and the NDT software to 38400.
This indicates a normal connection.
Note When using the wireless USB adapter, if the serial port repeatedly disconnects, check
that the battery is fully charged.
If desired, your Niton XL2 analyzer has the capability to download and store each reading to
the PC in real time to a file of your choice. To enable this feature, you must do the following:
• Your Niton analyzer must be turned on and connected to the PC. See
With Your
PC.
• The NDTr program module must be running and connected to your analyzer. See
Operating Your Analyzer
• The Download icon in the NDTr toolbar must be selected.
Figure 66. Live Download Icon
The file created is in a format readable by the NDT program module, has an extension of
.ndt, and looks identical to a file of manually downloaded readings - see
It can also create a simultaneous .CSV file.
Please note the following:
1. When the instrument is unplugged, selecting the Download icon does nothing.
Remotely.
Simultaneous Save as CSV File
Using Your Analyzer
Standard
Download.
.