Operation
Settings
Download Data To USB
3-128
42iQTL Instruction Manual
The Download Data To USB Drive screen allows the user to
download/upload data to/from the USB drive.
Drive
Home Screen>Settings>USB Drive>Download Data To USB Drive
The Download Data to USB Drive screen contains the following
information:
Download Health Check Report: Includes status and alarms, preventive
maintenance, and maintenance history.
Download Entire Data Log: Includes the entire data log (from data
logging).
Download Service Log: Includes a complete listing of data for all
variables. This is set at the factory.
Download System Log: Consists of system log text files, which include a
listing of system errors.
Download Calibration History: Includes the data in the calibration
history screen.
Download Configuration Data Backup to USB: Allows the user to
download the configuration file from the instrument to the USB.
Upload Configuration Data Restore from USB: Allows the user to upload
the configuration files from the USB to the instrument.
Restore: Allows the user to upload the configuration files from the USB
to the instrument.
Download All Data: Downloads all reports, logs, histories, and backup
information.