Table 5-1:
Special mode and process variable updates
Always polled and updated
•
Mass flow
•
Volume flow
•
Gas standard volume flow
•
Density
•
Temperature
•
Drive gain
•
LPO amplitude
•
Status [contains Event 1 and Event
2 (basic events)]
•
Mass total
•
Volume total
•
Live zero
•
Gas standard volume total
•
Temperature-corrected volume
total
•
Temperature-corrected density
•
Temperature-corrected volume
flow
•
Batch-weighted average tempera-
ture
•
Batch-weighted average density
5.1.2
Configure Response Time
ProLink III
Field Communicator Not available
Overview
Response Time is used to apply a different algorithm to the calculation of process variables
from the raw process data.
Restriction
Response Time is available only on systems with the enhanced core processor.
Procedure
Set Response Time as desired.
Configuration and Use Manual
Updated only when the petroleum
measurement application is disa-
bled
•
RPO amplitude
•
Core input voltage
•
Mass inventory
•
Volume inventory
•
Gas standard volume inventory
Device Tools > Configuration > Process Measurement > Response > Response Time
Configure device options and preferences
Never updated
All other process variables and calibra-
tion data. They retain the values held
at the time you enabled Special mode.
65