Figure 11-1:
D3 or D3 and D4 density calibration using ProLink III
11.9.2
Perform a D3 or D3 and D4 density calibration using the
Field Communicator
See the following flowchart.
Configuration and Use Manual
Close shutoff valve
downstream from sensor
D3 Calibration
Fill sensor with D3 fluid
Device Tools >
Calibration >
Density Calibration >
Density Calibration - Point 3
Enter density of D3 fluid
Start Calibration
Close
D4 Calibration
Fill sensor with D4 fluid
Device Tools >
Calibration >
Density Calibration >
Density Calibration - Point 4
Enter density of D4 fluid
Start Calibration
Close
Done
Measurement support