Check the core processor LED

The core processor has an LED that indicates different meter conditions.
1.
2.
3.
Postrequisites
To return to normal operation, replace the core processor lid.
Important
When reassembling the meter components, be sure to grease all O-rings.
12.31.1

Core processor LED states

Table 12-8:
Standard core processor LED states
LED state
1 flash per second (ON 25%, OFF
75%)
1 flash per second (ON 75%, OFF
25%)
Solid ON
3 rapid flashes, followed by
pause
4 flashes per second
194
Plug the terminal blocks into the terminal board.
Replace the lid on the sensor junction box.
Maintain power to the transmitter.
Remove the core processor lid. The core processor is intrinsically safe and can be
opened in all environments.
Check the state of the core processor LED.
Description
Normal operation
Slug flow (two-phase flow)
Zero or calibration in progress
Core processor receiving be-
tween 11.5 and 5 volts
Sensor not recognized
Improper configuration
Broken pin between sensor and
core processor
Fault condition
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Recommended actions
No action required.
See
Section
12.27.
No action required.
Check power supply to transmitter.
Check wiring between transmitter and sensor.
Check sensor characterization parameters.
The meter requires factory service.
Check alert status.