Emerson Micro Motion 2500 Configuration And Use Manual page 76

Transmitters with configurable input/outputs
Hide thumbs Also See for Micro Motion 2500:
Table of Contents

Advertisement

Configure device options and preferences
Status alerts and options for Status Alert Severity
Table 5-2:
Status alerts and Status Alert Severity
Alert code
Status message
A001
EEPROM Error
A002
RAM Error
A003
No Sensor Response
A004
Temperature Overrange
A005
Mass Flow Rate Overrange
A006
Characterization Required
A008
Density Overrange
A009
Transmitter Initializing/
Warming Up
A010
Calibration Failure
A011
Zero Calibration Failed:
Low
A012
Zero Calibration Failed:
High
A013
Zero Calibration Failed:
Unstable
A014
Transmitter Failure
A016
Sensor RTD Failure
A017
T-Series RTD Failure
A018
EEPROM Error (Transmit-
ter)
A019
RAM Error (Transmitter)
A020
No Flow Cal Value
A021
Incorrect Sensor Type (K1)
A022
Configuration Database
Corrupt (Core Processor)
68
Option
Description
Informa-
Actions when fault is detected:
tional
The alert is posted to the Alert List.
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
The status LED (if available) returns to green and may or may not flash.
Ignore
No action
Default severity Notes
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Applies only to flowmeters with the
standard core processor.
Configurable?
No
No
Yes
No
Yes
Yes
Yes
Yes
No
Yes
Yes
Yes
No
Yes
Yes
No
No
Yes
No
No

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents