Emerson Micro Motion 2500 Configuration And Use Manual page 176

Transmitters with configurable input/outputs
Hide thumbs Also See for Micro Motion 2500:
Table of Contents

Advertisement

Troubleshooting
Alert num-
ber
Alert title
A104
Calibration in Pro-
gress
A105
Two-Phase Flow
A106
Burst Mode Enabled
A107
Power Reset Occur-
red
A108
Basic Event 1 On
A109
Basic Event 2 On
A110
Frequency Output
Saturated
A111
Frequency Output
Fixed
168
Possible cause
A calibration procedure is in proc-
ess.
The line density is outside the user-
defined two-phase flow limits.
HART burst mode is enabled.
The transmitter has been restarted.
The process has triggered Basic
Event 1.
The process has triggered Basic
Event 2.
The calculated Frequency Output is
outside the configured range.
The Frequency Output has been
configured to send a constant val-
ue.
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Recommended actions
Allow the procedure to complete.
For zero calibration, you may abort the
calibration, set Zero Time to a lower val-
ue, and restart the calibration.
Check for two-phase flow.
Check the live density reading against
the upper and lower two-phase flow
limit settings.
No action required.
If desired, you can set Alert Severity
Level to Ignore.
No action required.
If desired, you can set Alert Severity
Level to Ignore.
No action required.
Review event configuration if you be-
lieve the event was triggered errone-
ously.
No action required.
Review event configuration if you be-
lieve the event was triggered errone-
ously.
Check the Frequency Output scaling.
Check process conditions. Actual condi-
tions may be outside the normal condi-
tions for which the output is config-
ured.
Check for air in the flow tubes, tubes
not filled, foreign material in the tubes,
coating in the tubes, or other process
problems.
Verify that the measurement units are
configured correctly for your applica-
tion.
Purge the sensor tubes.
Cycling power to the meter or restart-
ing totalizers will restore the Frequency
Output to normal operation.
Check whether the output is in loop test
mode. If it is, unfix the output.
Check whether the output has been set
to a constant value via digital communi-
cation.