Emerson Micro Motion 2500 Configuration And Use Manual page 77

Transmitters with configurable input/outputs
Hide thumbs Also See for Micro Motion 2500:
Table of Contents

Advertisement

Table 5-2:
Status alerts and Status Alert Severity
Alert code
Status message
A023
Internal Totals Corrupt
(Core Processor)
A024
Program Corrupt (Core
Processor)
A025
Boot Sector Fault (Core
Processor)
A026
Sensor/Transmitter Com-
munications Failure
A027
Security Breach
A028
Core Processor Write Fail-
ure
A031
Low Power
A032
Meter Verification in Pro-
gress: Outputs to Fault
A033
Insufficient Right/Left Pick-
off Signal
A034
Meter Verification Failed
A035
Meter Verification Aborted Fault
A100
mA Output 1 Saturated
A101
mA Output 1 Fixed
A102
Drive Overrange
A103
Data Loss Possible (Totals
and Inventories)
A104
Calibration in Progress
A105
Slug Flow
A106
Burst Mode Enabled
A107
Power Reset Occurred
Configuration and Use Manual
(continued)
Default severity Notes
Fault
Applies only to flowmeters with the
standard core processor.
Fault
Applies only to flowmeters with the
standard core processor.
Fault
Applies only to flowmeters with the
standard core processor.
Fault
Fault
Fault
Fault
Applies only to flowmeters with the
enhanced core processor.
Varies
Applies only to transmitters with
Smart Meter Verification.
If outputs are set to Last Measured
Value, severity is Info. If outputs are
set to Fault, severity is Fault.
Fault
Applies only to flowmeters with the
enhanced core processor.
Fault
Applies only to transmitters with
Smart Meter Verification.
Applies only to transmitters with
Smart Meter Verification.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Informational
Applies only to flowmeters with the
enhanced core processor.
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Normal transmitter behavior; oc-
curs after every power cycle.
Configure device options and preferences
Configurable?
No
No
No
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
69