Transmitter status reported by LED

Table 2-1:
LED state
Solid green
Flashing yellow
Solid yellow
Solid red
2.3

Make a startup connection to the transmitter

To configure the transmitter, you must have an active connection from a communications
tool. Follow this procedure to make your first connection to the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in
the appropriate appendix. Use the default communications parameters shown in the
appendix.
Communications tool
ProLink III
Field Communicator
6
Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions. During the power-up sequence, Alert A009 is active.
This alert should clear automatically when the power-up sequence is complete.
Check the status LED on the transmitter.
View and acknowledge status alerts
Status LED states
Alarm condition
No alarm
No alarm
Low-severity alarm
High-severity alarm
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Description
Normal operation
Zero calibration procedure is in progress
Loop test is in progress
Alarm condition that will not cause measure-
ment error (outputs continue to report proc-
ess data)
Alarm condition that will cause measurement
error (outputs in fault)
Connection type to use
Modbus/RS-485
HART
Instructions
Appendix A
Appendix B