Operate the transmitter with the Weights & Measures application
Table 10-5:
Available actions when Approvals=NTEP
Function
Connecting from ProLink II, ProLink III, the
Field Communicator, or a host
Configuration
Calibration
Smart Meter Verification
External pressure and
temperature data
Testing and trimming out-
puts
Totalizers
Inventories
Sensor simulation
136
Unsecured
All supported connec-
tion types are availa-
ble.
Reading
Allowed
Changing
Allowed
Zero
Allowed
Density
Allowed
Temperature
Allowed
Outputs set to Continue
Allowed
Measuring
Outputs set to Fault
Allowed
Outputs set to Last
Allowed
Measured Value
Retrieved by polling
Allowed
Written by Modbus or
Allowed
HART host
mA output loop test
Allowed
mA output trim
Allowed
Frequency output loop
Not allowed
test
Discrete output loop
Allowed
test
Resetting
Allowed. If any single
totalizer is reset, all
other totalizers are re-
set automatically.
Starting
Not allowed
Stopping
N/A
Resetting
Not allowed
Allowed
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Transmitter status
Secured
Physical security prevents access to the
RS-485 terminals, so neither RS-485
connections nor service port connec-
tions are available.HART/Bell 202 con-
nections can be used to read data from
the transmitter, but software security
prevents writing to the transmitter.
Allowed
Not allowed
Not allowed
Not allowed
Not allowed
Allowed
Allowed
Not allowed
Allowed
Not allowed
Not allowed
Not allowed
Not allowed
Not allowed
May or may not be allowed, depending
on configuration. Can be performed on-
ly if flow rate is zero. If any single totaliz-
er is reset, all other totalizers are reset
automatically.
N/A
Not allowed
Not allowed
Not allowed