View Test Data - Emerson Micro Motion 2500 Configuration And Use Manual

Transmitters with configurable input/outputs
Hide thumbs Also See for Micro Motion 2500:
Table of Contents

Advertisement

Measurement support
Run an SMV test using the Field Communicator
1.
2.
3.
4.
Postrequisites
View the test results and take any appropriate actions.
11.2.4

View test data

You can view the results of the current test. You can also view results from previous tests.
Important
You can view previous test results and see detailed test reports only if SMV is licensed.
The transmitter stores the following information about the previous twenty SMV tests:
In addition, ProLink III provides a detailed test reporting and analysis framework. This
information is stored on the PC where ProLink III is installed for tests that were run only on
that PC. It includes:
144
Navigate to the Smart Meter Verification menu:
Overview > Shortcuts > Meter Verification
Service Tools > Maintenance > Routine Maintenance > Meter Verification
Choose Manual Verification.
Choose Start.
Set output behavior as desired, and press OK if prompted.
Option
Description
Continue Measuring
During the test, all outputs will continue to report their assigned proc-
ess variable. The test will run for approximately 90 seconds.
Outputs Held at Last
During the test, all outputs will report the last measured value of their
Value
assigned process variable. The test will run for approximately
140 seconds.
Outputs Held at Fault During the test, all outputs will go to their configured fault action. The
test will run for approximately 140 seconds.
Test progress is displayed on the screen.
Powered-on hours at the time of the test.
Test result (Pass, Fail, Abort).
Abort code, if applicable.
Timestamp from the PC clock
Current flowmeter identification data
Current flow and density configuration parameters
Current zero values
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs