C H A P T E R
1 4
-
A T
To program the test:
1. Enter the voltage and frequency parameters of the test signal.
(If in doubt, refer to Chapter 7 - 'Test Conditions', which lists
recommended test conditions).
2. Select the integration time you require
'Medium' is the default setting. 'Long' will give more stable readings (for
tighter limits) at the expense of an increased test time, and 'Short' will test
at the maximum speed, but may give a slightly noisier reading. See 7.3
Explanation of Integration.
3. Enter the names for the energized and measured terminals for the X
measurement.
4. Similarly configure Y terminals.
5. When required the "Optional 2nd Energized Hi" allows an additional terminal
connection to the energized high node.
6. Set (X/Y) limits.
+/- Enter a nominal value and then the limits as negative and positive
values.
>< Enter minimum and maximum values.
> Enter just a minimum value.
Note: The Measure button may be used here as outlined in section 14.5.9.
7. Since components on the test fixture may scale measured readings the
"Compensate for Test Fixture Scaling" allows the test to account for this.
8. Select OK. The test specification will then be displayed in the program
window.
AT5600 User Manual 98-119 issue 14